C-SCOPE CS7 Manual de servicio Pagina 62

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 83
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 61
GE Medical Systems SIGNA PROFILE CONFORMANCE STATEMENT
REV 2.7 sm 2303347
doc ver 2.7 62
Appendix A
IDO Definitions
Enclosed is a listing of only the Optional/Conditional modules/attributes used by this implementation for CT/MR/SC
and Standalone Overlay IODs.
NOTE:
All other modules which are mandatory and consist of only type 1 elements and attributes which are type 1 pre
DICOM standard are supported by this implementation but are not explicitly listed here.
A.1
CT Specific IOD Definition
A.1.1
CT Image IOD Modules
Entity Name Module Name Reference Usage
Patient Patient A.5.1 M
Study General Study A.5.2 M
Patient Study A.5.3 U
Series General Series A.5.4 M
Frame of Reference Frame of Reference A.5.12 M
Equipment General Equipment A.5.5 M
Image General Image A.5.6 M
Image Plane A.5.7 M
Image Pixel Only the type 1 elements are sent. M
Contrast/Bolus A.5.8 C
CT Image A.1.2 M
SOP Common A.5.9 M
A.1.2
CT Image Module
Attribute Name Tag Type Notes
Rescale Intercept (0028,1052) 1 -1024 or 0 sent
Rescale Slope (0028,1053) 1 Value of 1 is sent
KVP (0018,0060) 2 Value always sent.
Acquisition Number (0020,0012) 2 Value always sent.
Date Collection Diameter (0018,0090) 3 Value always sent.
Reconstruction Diameter (0018,1100) 3 Value sent for all images except scouts.
Distance Source to Detector (0018,1110) 3 Value always sent.
Distance Source to Patient (0018,1111) 3 Value always sent.
Gantry / Detector Tilt (0018,1120) 3 Value always sent.
Table Height (0018,1130) 3 Value always sent.
Rotation Direction (0018,1140) 3 Not sent for scout or axial
Exposure Time (0018,1150) 3 Value always sent.
Xray Tube Current (0018,1151) 3 Value always sent.
Exposure (0018,1152) 3 Value always sent.
Filter Type (0018,1160) 3 Sent.
Defined terms:
BODY FILTER
BOWTIE FLAT FILT
FLAT FILTER
LARGE BOWTIE FIL
Focal Spot (0018,1190) 3 Sent. Fixed value of 0.7 or 1.2
Convolution Kernel (0018,1210) 3 Not sent for scouts.